| Model Number |
3543407MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Pain (1994); No Code Available (3191)
|
| Date of Event |
08/31/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture, breast pain.(b)(4).Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient who underwent a breast augmentation with two textured mentor memorygel breast implants (a revision procedure with 340cc on the left and a primary procedure with 350cc on the right) has experienced bilateral rupture of implants and bilateral breast pain postoperatively.As a result, the patient is scheduled to undergo explantation and replacement with unspecified mentor breast implants on (b)(6) 2020.This medwatch report is for the left-sided implant.
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received and completed the evaluation of the suspect medical device.Device evaluation summary: during a visual evaluation of the product, parallel lines of shell wear were observed on the anterior aspect.Leak testing was performed, according with mentor procedure, and it revealed a tear within the shell wear measuring approximately less than 0.1 cm.The evaluation determined that the loss of shell integrity is consistent with a rupture caused by wear.Shell wear suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.Most women undergoing augmentation or reconstruction with a mammary prosthesis will experience some pain postoperatively.While pain normally subsides in most women as they heal from surgery, it can become a chronic problem in other women.Chronic pain can be associated with a variety of factors.Surgeons should instruct their patients to inform them if there is significant pain or if pain persists.Pain is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information is consistently analyzed and monitored by mentor quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Mentor received additional event information that the patient also experienced bilateral capsular contracture, baker grade unknown.As a result, the patient underwent replacement with mentor memorygel breast implant, 350cc on the left (catalog # 3503501bc, left lot-serial # (b)(6)) and 400cc on the right (catalog # 3504001bc, right lot-serial # (b)(6)) on (b)(6) 2020.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has completed the re-evaluation of the suspect medical device.The device evaluation summary was updated to include: mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination manufacturer¿s reference number: (b)(4).
|
| |
|