• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550350-23
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported the device was discarded.Investigation is not yet complete.A follow-up will be submitted with all relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a mildly calcified lesion in the moderately tortuous proximal left circumflex (plcx) coronary artery.The 3.50x23 mm xience sierra stent was attempted but could not cross the lesion due to the tortuous anatomy.The device was strongly pushed, causing the stent edge to flare.There was resistance met with the guiding catheter during removal.Another stent was used to complete the procedure.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Product performance engineering reviewed the incident information; however, the product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.There was no damage noted to the stent delivery system (sds) during the inspection prior to use which suggests a product quality issue with respect to manufacture, design or labeling did not contribute to the reported difficulties.The investigation determined the reported failure to advance, difficulty to remove and stent damage appear to be related to operational circumstances of the procedure.Based on the reported information, during advancement the sierra sds encountered resistance with the mildly calcified and moderately tortuous anatomy resulting in the failure to advance.Mishandling of the device or the device being strongly pushed against resistance resulted in the reported stent damage and difficulty to remove from the guiding catheter.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10570909
Report Number2024168-2020-07933
Device Sequence Number19616465
Product Code NIQ
UDI-Device Identifier08717648227431
UDI-Public08717648227431
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/23/2023
Device Model Number1550350-23
Device Catalogue Number1550350-23
Device Lot Number0061041
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/02/2020
Supplement Date Received by Manufacturer09/23/2020
Initial Report FDA Received Date09/22/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-