| Model Number |
MMT-722LNAP |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/07/2013
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump passed the displacement test.Unit received with motor error alarm during the basic occlusion test due to loose/flush drive support disk.Unable to perform prime/compromised force sensor system test, excessive no delivery test and occlusion test or verify no excessive no delivery alarm due to motor error alarm.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had no delivery alarm and insulin was exited.Customer experienced high blood glucose level.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|