Model Number 8886848813 |
Device Problem
Device Slipped (1584)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/31/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic cholecystectomy, while the surgeon was trying to clip bile duct, the clips slipped off from the tissue.Another ligation clip product was used to complete the case.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the clip did not hold tightly to the vessel and detached.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
|
|