Model Number N/A
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 09/04/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) was dispatched to evaluate the iabp unit.The shaft lock was replaced.The stm completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.(b)(6).
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) shaft collar would not lock.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h10.After further review, an initial emdr for this complaint was incorrectly submitted.This is a non-reportable complaint with a classification of "other", making it non-reportable to the fda.As a result, no follow up emdr is necessary.Please cancel in your database.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) shaft collar would not lock.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10573661
MDR Text Key208156961
Report Number2249723-2020-01542
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received01/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-