Model Number 25VAVGJ-515
Medical Device Problem Codes Fracture (1260); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
On (b)(6) 2020, a 25mm sjm masters series valsalva aortic valved graft was selected for implant in the patient.During the aortic root replacement procedure, the physician used the valve rotator recommended in the ifu, but one of the leaflets was damaged by rotating the discs.One of the leaflets disconnected from the hinge, the valve was removed from the patient, and a carbomedics valve was implanted instead.The patient remained stable throughout the procedure, but the procedure was extended by 30-45 minutes.The patient is currently stable.
 
Additional Manufacturer Narrative
The reported event of a dislodged leaflet and damaged leaflet was confirmed.Two leaflets were received, one of which was fractured.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.While the root cause of the leaflet dislodgement could not be conclusively determined, there was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflets or orifice damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALSALVA AORTIC VALVED GRAFT
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key10574969
Report Number2648612-2020-00097
Device Sequence Number3074388
Product Code LWQ
UDI-Device Identifier05414734009492
UDI-Public05414734009492
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number25VAVGJ-515
Device Catalogue Number25VAVGJ-515
Device Lot Number7365362
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/31/2020
Supplement Date Received by Manufacturer11/09/2020
Initial Report FDA Received Date09/23/2020
Supplement Report FDA Received Date12/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-