Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00227, 0001032347-2020-00456, 0001032347-2020-00458.Implantation date was in 2010.Tmj system left fossa component, medium, part# 24-6561, lot# 080260 tmj system right fossa component, medium, part# 24-6560-int, lot# 725240 tmj system left standard mandibular component 55mm / 12 hole, part# 24-6556, lot# 205450 tmj system right standard offset mandibular component 50mm / 9 hole, part# 24-6650-int, lot# 489880.Unknown screws, part# ni, lot# ni.(b)(6).User facility is foreign; therefore, a facility medwatch report will not be available.Report source: (b)(6).
|
|
This follow-up report is being submitted to relay additional information.The complaint is considered confirmed because a revision surgery was reported.No product was returned; therefore, no functional tests or inspections could be performed.No x-rays, scans, pictures, or physician's reports were provided.The non-conformance database was reviewed for the mandible component; no non-conformances were found.There are no indications of manufacturing defects.This is the only complaint for this item# 24-6556, lot# 205450.For all non-custom tmj mandibular implants in the previous one year (from the notification date) regarding revision due to ankyloses, there is a complaint rate of 0.07%, which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is due to a patient condition.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|