• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284002
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the outflow performance failed.Per service report, the reported problem was not duplicated during evaluation; therefore, this complaint cannot be confirmed.Unrelated to the reported problem, it was found that the fill led indicator was not working.The bezel/touch panel assembly was replaced to resolve the issue.Both pressure tips assemblies were also replaced.After repair, the device was found to be working according to the specifications.As the reported problem was not confirmed, a root cause for the issue that was experienced by the user cannot be determined.Also, we cannot discern the root cause of the identified problem.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 03/27/2019 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Event Description
It was reported by j&j employee via complaint submission tool, that during internal service activities, such as calibration, a fms vue fluid management system, did not fulfill required acceptance criteria due to outflow performance failed during commissioning. no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10576607
MDR Text Key208336469
Report Number1221934-2020-02724
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284002
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2020
Initial Date Manufacturer Received 09/22/2020
Initial Date FDA Received09/23/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-