SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE Back to Search Results
Catalog Number 72200873
Device Problem Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Date 08/27/2020
Event Type  malfunction  
Event Description
It was reported that the controller was overheating and had a burning smell that was being emitted.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3,h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of overheating and burnt smell was confirmed.Product failed functional testing with a short circuit error.Cause of errors is a shorted electronic component on the main digital pcb.Burnt odor was caused by a burned resistor on main pcb.Product passed functional testing with a known good main pcb installed.The complaint investigation has concluded the cause of the failure to be defective electronic components on the main digital pcb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10577004
MDR Text Key208390190
Report Number1643264-2020-01271
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Initial Date Manufacturer Received 09/01/2020
Initial Date FDA Received09/23/2020
Supplement Dates Manufacturer Received10/12/2020
Supplement Dates FDA Received10/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-