It was reported that the controller was overheating and had a burning smell that was being emitted.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H3,h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of overheating and burnt smell was confirmed.Product failed functional testing with a short circuit error.Cause of errors is a shorted electronic component on the main digital pcb.Burnt odor was caused by a burned resistor on main pcb.Product passed functional testing with a known good main pcb installed.The complaint investigation has concluded the cause of the failure to be defective electronic components on the main digital pcb.
|