• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE 3-LUMEN SPHINCTEROTOME V; SINGLE USE 2-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0730
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc).The exact cause has been under investigation.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event or Problem Description
During an endoscopic sphincterotomy (est), the subject device was used.The cutting wire of the subject device broke.The intended procedure was completed with another device.There was no patient injury reported.
 
Additional Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The cutting wire of the subject device was broken.The broken section of the cutting wire was melted and burned.As a measurement result of the outer diameter of the cutting wire, there was no abnormality.As a measurement result of the length of the cutting wire and the cutting wire coating, there was no abnormality.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, omsc presumes that the event occurred due to the following mechanism.The cutting wire contacted to the distal part of the endoscope since the rear end of the cutting wire was not extended from the distal end of the endoscope.The subject device was activated, and it caused spark.A part of the cutting wire became extremely hot, resulting in breakage.The above device handling has warned in the instruction manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10578539
Report Number8010047-2020-06823
Device Sequence Number2627859
Product Code KNS
UDI-Device Identifier04953170382642
UDI-Public04953170382642
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V411M-0730
Device Lot Number04K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/04/2020
Supplement Date Received by Manufacturer09/28/2020
Initial Report FDA Received Date09/24/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-