MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 42

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; TURON PEGGED GLENOID IMPLANT, SIZE 42 Back to Search Results
Model Number 520-01-242
Health Effect - Clinical Codes Pain (1994); Limited Mobility Of The Implanted Joint (2671)
Date of Event 08/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - patient had pain and limited movement.Ct scan and x-ray indicated wear and joint loss.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as ware and loose joint.The previous surgery and the surgery detailed in this event occurred 7 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to ware and loose joint.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
TURON PEGGED GLENOID IMPLANT, SIZE 42
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key10580159
Report Number1644408-2020-00882
Device Sequence Number16921698
Product Code KWS
UDI-Device Identifier00888912024952
UDI-Public(01)00888912024952
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123982
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/28/2017
Device Model Number520-01-242
Device Catalogue Number520-01-242
Device Lot Number191G1045
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/27/2020
Supplement Date Received by Manufacturer10/21/2020
Initial Report FDA Received Date09/24/2020
Supplement Report FDA Received Date11/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-000 LOT 878C1222; 520-01-012 LOT 456G1060; 520-50-126 LOT 938C1006; 520-00-000 LOT 878C1222; 520-01-012 LOT 456G1060; 520-50-126 LOT 938C1006
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-