|
Multiple attempts were made to obtain additional information from the customer regarding the event (including device serial number); however, no additional information was provided.Manufacturer's investigation conclusion: the reported events of system controller faults and damage to the white power lead of the system controller were not confirmed.The system controller was not returned for evaluation and no log files or photos were submitted for review.Multiple requests were made to obtain additional information (including if the system controller faults resolved with the controller exchange, if the controller will be returned for evaluation, and the serial number of the controller); however, no response was received.The root cause of the reported events could not be conclusively determined through this analysis.Device history records could not be reviewed as the serial number of the product is unknown.Heartmate ii patient handbook under section 5 ¿alarms and troubleshooting¿ and heartmate ii instructions for use, under section 7 ¿alarms and troubleshooting¿ cover all alarms (visual and audible), including the controller fault alarm conditions, and the actions to take if the alarms cannot be resolved.Heartmate ii patient handbook section 6 "caring for the equipment" describes how to care for and clean all equipment, including the system controller power cables.Section 10, entitled ¿safety checklists¿, provides checklists to assist the patient in performing routine maintenance of heartmate ii lvad, including inspecting the system controller power cables for damage.This section also informs the user to replace any equipment or system component that appears damaged or worn.Heartmate ii patient handbook section 2 "how your heart pump works" and heartmate ii instructions for use (ifu) section 2 "system operations" explain how maintain the backup system controller and how to replace the running system controller with the backup controller.Heartmate ii patient handbook and heartmate ii instructions for use (ifu) caution users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|