|
A discordant, falsely elevated prothrombin time international normalized ratio (inr) result was obtained on a patient sample on a sysmex cs-2100i system using thromborel s reagent.Due to the high result, the sample was rerun for inr using the same system and same reagent, resulting lower.The same sample was then repeated for inr on a sysmex cs-5100 system using thromborel s reagent, resulting lower.The initial discordant result obtained on the sysmex cs-2100i system was auto-validated and reported to the physician(s).The customer informed the physician(s) to disregard the falsely elevated inr result and requested that another sample from the patient be obtained and run for inr.The patient was redrawn and the sample was run for inr, resulting lower and matching the result previously obtained on the sysmex cs-5100 system.This result was reported, as the correct result, to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated prothrombin time international normalized ratio (inr) results.
|