| Model Number |
60005 |
| Medical Device Problem Code |
Misfocusing (1401)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
08/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Doctor entered the incorrect target for the light delivery device.The light delivery device adjusted the lens to the target as specified by the doctor, resulting in undesired refractive outcome for the patient.Device history record for the lens was reviewed.No issues were noted.The explanted lens has been returned and is currently being evaluated.
|
| |
|
Event or Problem Description
|
|
Doctor entered the incorrect target for the light delivery device.The light delivery device adjusted the lens to the target as specified by the doctor, resulting in undesired refractive outcome for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The explanted lens was returned.The lens was cut into several fragments, with some of the cuts going through the center of the optic.Due to the extent of damage to the lens, additional testing could not be performed.
|
| |
|