Device manufactured between march 30, 2020 to april 02, 2020.The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed and no leak or air bubbles were observed.System level testing was performed using sterile water and 70 % dextrose attached to the exactamix main module and no air bubbles or leaks were observed from the sample.Accuracy tests were performed and found to be within the acceptable limits.No issues were observed during the testing of the valve set.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|