SMITH & NEPHEW, INC. MI TRIAL FEM HEAD 32 -3; PROS,KNEE,FEM,SEMICONSTRAINED,CEMEN,MET/POLY
|
Back to Search Results |
|
| Model Number |
71369714 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/31/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the femoral trial head confirms the device has multiple scratches, burrs and deep gouges in the plastics the device show extreme signs of wear/usage.There is no lot number available.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that upon inspection the plastic ring inside the trial was torn, the outside was extremely scratched up and worn out.No case involved.
|
| |
|
Search Alerts/Recalls
|
|
|