| Model Number |
CTS02 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.This report is to follow-up medwatch # (b)(4).
|
| |
|
Event or Problem Description
|
|
Procedure performed: laparoscopic peritoneal dialysis.Event description: received via mail, medwatch uf/importer report # (b)(4)."while inserting the trocar, it bent." additional information received via email on 28aug2020 from user facility: the location of the bend was on the mid shaft of trocar.The obturator was inside the cannula at the time of the incident.It is unknown if there was difficulty with the insertion.The whole product was replaced.It is unknown how the case was completed after the event occurred.The unit is available for return.Additional information received via phone on 17sep2020 from account manager: the facility is returning a obturator from a ctf03 of lot 1384278 and a cannula from a cts02 of lot 1383546.They were both used in the same case.The rep was not able to confirm if the products were used together in the case.Intervention: the whole product was replaced.Patient status: no adverse event reported.Reported only as product problem.
|
| |
|
Event or Problem Description
|
|
The event unit contained components from two different models: component 1 of 2 (2027111 - 2020 - 00546), component 2 of 2 (2027111 - 2020 - 00565).Event description: received via mail, medwatch uf/importer report # (b)(4)."while inserting the trocar, it bent." event occurred in july 2020.Additional information received via phone on 27aug2020 from user facility: the event date is 15jul2020.Additional information received via email on 28aug2020 from user facility: the location of the bend was on the mid shaft of trocar.The obturator was inside the cannula at the time of the incident.It is unknown if there was difficulty with the insertion.The whole product was replaced.It is unknown how the case was completed after the event occurred.The unit is available for return.Additional information received via phone on 17sep2020 from applied medical account manager i: the facility is returning a obturator from a ctf03 of lot 1384278 and a cannula from a cts02 of lot 1383546.They were both used in the same case.The rep was not able to confirm if the products were used together in the case.Intervention: the whole product was replaced.Patient status: no adverse event reported.Reported only as product problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of a bent cannula.The wall thickness of the cannula shaft was measured and met specifications.Based on the condition of the returned unit and the description of the event, it is likely that the cannula bent due to a high insertion force that was applied to the unit during insertion.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.A bent cannula is not considered to be reportable as it is unlikely to cause or contribute to death or serious injury.This report is to follow-up medwatch #: (b)(4).
|
| |
|