H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of difficulty activating the safety mechanism was confirmed.The product returned for evaluation was a one 22ga x 0.75" huber plus safety infusion set.The returned product sample was evaluated and the needle was observed to be bent at the exit site from the needle housing.The following observations were noted during the sample evaluation: ¿ usage residue was seen on the sample which proved that the product had experienced at least some use ¿ the needle tip was barbed suggesting contact between the needle and port base the sample was received with the safety partially engaged.The safety was easily fully engaged during preliminary evaluation; however, the needle damage suggested that needle deformation may have complicated safety activation during device removal from the port.Force applied at an angle to the needle axis, or if the needle is not inserted perpendicular into the port septum, can lead to bending of the needle shaft.It is advised to verify that the needle length is correct based on port reservoir depth, tissue thickness and the thickness of any dressing beneath the bend of the needle; if too long, needle and/or port may be damaged at insertion.Additionally, avoid excessive manipulation once the needle is in the port.No potential damage related to the manufacturing process was noted on the complaint sample.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
|