• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CONQUEST FN SCREW EXTENSION ROD; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CONQUEST FN SCREW EXTENSION ROD; DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT Back to Search Results
Model Number 71179216
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/31/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery outside of the patient, the screwdriver and the extension rod both got broken, as well as a missing piece that holds the screw in both positions.There was no delay nor injury reported.A s&n backup device was available to finished the surgery.
 
Additional Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST FN SCREW EXTENSION ROD
Common Device Name
DEVICE, FIXATION, PROXIMAL FEMORAL, IMPLANT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10584927
Report Number1020279-2020-04859
Device Sequence Number11520117
Product Code JDO
UDI-Device Identifier00885556683620
UDI-Public00885556683620
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152686
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71179216
Device Catalogue Number71179216
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 08/31/2020
Supplement Date Received by Manufacturer03/28/2021
Initial Report FDA Received Date09/25/2020
Supplement Report FDA Received Date04/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-