|
No damages or irregularities were found with the concerto pusher.The implant coil was found broken from the detach element with the polypropylene filament also broken.The broken implant coil was not returned for analysis.The concerto implant coil could not be used for resistance testing due to its damaged condition.No damages or irregularities were found with the stc-18 micro catheter hub.The stc-18 catheter body was found kinked at ~29.6cm, and found flattened at ~4.2cm from the distal end.The stc-18 has an inner diameter of 0.021¿ (per boston scientific website) which is compatible for use with the concerto coil.An in-house axium prime (model: apb-4-6-hx-ss lot: a280237) was then used for resistance testing.The coil was inserted into the catheter hub, through the lumen and out the distal end with no resistance encountered.All other subassemblies appeared to be normal and no other anomalies were observed.Based on the device analysis and reported information, the customer¿s report of ¿coil resistance/stuck in catheter¿ could not be co nfirmed and the root cause could not be determined.Possible causes for coil resistance in catheter are, but not limited to severely tortuous patient anatomy, failure to hydrate the concerto coil prior to use and lack of continuous flush during delivery.Customer reported vessel tortuosity as moderate.It is possible that the kink and the flattened section of the catheter contributed towards the resistance, however, the cause of the damage could not be determined.The implant coil was found broken.It is likely the implant coil became broken when attempting to advance/retract the device against the reported resistance.Other possible causes of implant coil break are lack of hydration before procedure, user does not maintain continuous flush, tortuous anatomy, coil not retracted in a one-to-one motion with the implant pusher during repositioning, pushwire rotation, or incompatible catheter.As the implant coil was not returned for analysis, any contribution of the implant coil towards the resistance could not be determined.There is no indication that the event is related to a manufacturing issue.If information is provided in the future, a supplemental report will be issued.
|