Information was received from a healthcare provider via a manufacturer representative regarding a patient with olif therapy.It was reported that, during the procedure cage went in and while pulling out it was noticed that the inserter was broken, the broken piece was also retrieved there was no fragment left or contact with the patient.There were no symptoms reported.There were no further complications reported regarding the event.
|
|
H3: product analysis#: (b)(4); lot# pr11c017; visual and optical examination confirmed one of the prongs has broken off; the broken piece is missing and not returned for analysis.Fracture initiation appears to be located at the base of the prong, consistent with bend stress overload.Optical examination of the fracture surface reveals a fairly brittle fracture, with no indication of fatigue.This type of damage is consistent with bend stress overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|