Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed that the long bipolar grasper instrument had damaged insulation to the conductor wire at the distal idler pulley.Material appeared to be removed, exposing the bare wire.Material appeared to be missing, with the piece being approximately 0.03" x 0.03".The electrical continuity test still passed.No thermal damage was observed at the wrist assembly.The long bipolar grasper instruments are multiple-use electrosurgical endoscopic instruments with a grasping tip to be used in conjunction with the da vinci system and an external electrosurgical unit (esu).The instrument is designed to provide energy from the designated location on the instrument (the tip) to the planned anatomical location when used as intended.The energy is activated by pressing the designated pedal on the surgeon side console (ssc).A site history review was performed and no related complaints were found to the product.Image/video review: no image or video clip for the reported event was submitted for review.An instrument log review confirmed that the long bipolar grasper instrument (470400-10/n10200317 0006) was last used on (b)(6) 2020 on system (b)(4) and had 3 uses remaining.This complaint is being reported because a bipolar instrument with damaged conductor wire insulation could lead to inadvertent transmission to tissue other than intended via the exposed conductor wire.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.Follow-up was attempted, but the patient information was either unknown, unavailable, not provided, or not applicable.The expiration date is not applicable.Implant date is not applicable because the product is not implantable.Initial reporter is blank because it is unknown if the initial reporter submitted a report to the fda.Recall is not applicable.
|