BIOMET 3I CERTAIN LOCATOR ABUTMENT 3.4MM(D) X 4MM(H) Back to Search Results
Model Number IMLOA004
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient identifier unknown/not provided.Age and date of birth unknown/not provided.Patient sex unknown/not provided.Weight unknown/not provided.Date of the event unknown/not provided.Product not returned.
 
Event or Problem Description
It was reported that the abutment fractured and was removed.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The complaint report was received, reviewed, and evaluated by the manufacturer complaint process in compliance with applicable fda and ous country regulations.Applicable lot history records were reviewed to evaluate whether any internal deviations, non-conformances, or problems occurred during the manufacture of the lot, which could cause or contribute to the reported complaint condition.A review of other manufacturers complaint files and related trending data for similar incidents was performed to evaluate whether other similar complaints have been received and if data suggests any adverse trends may have contributed to the complaint root cause.Investigation results / root cause or most-likely root-cause: as part of each complaint investigation, the manufacturer performs an analysis to determine the root-cause of the reported problem.Depending upon the information provided, a true root-cause may or may not be determined.In these cases, the most-likely cause of the reported issue would be determined.Other known issues that are inherent to the business operations, typically ones that are related to shipping, handling, customer service, or other, are internally tracked and continually trended for unusual increases.As a result of the above investigation, zest has concluded the following as the root-cause or most likely root-cause of the reported complaint condition: the reported complaint condition cannot be verified thread fracture during function.Most likely root cause of the failure mode / hazard : typically is the result from insufficient tightening torque at placement, overloading of the abutment in function, or not retightening the abutment at prescribed intervals.No findings are available since no physical evaluation was performed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN LOCATOR ABUTMENT 3.4MM(D) X 4MM(H)
Common Device Name
ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10592223
Report Number0001038806-2020-01455
Device Sequence Number13741194
Product Code DZE
UDI-Device Identifier00844868006343
UDI-Public(01)00844868006343(17)230824(10)1212372(241)IMLOA004
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K012911
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/24/2023
Device Model NumberIMLOA004
Device Catalogue NumberIMLOA004
Device Lot Number1212372
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/02/2020
Supplement Date Received by Manufacturer11/10/2020
Initial Report FDA Received Date09/28/2020
Supplement Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-