(b)(4).Date sent to the fda: 11/9/2020 additional information: d4, d10, h4, h6 a manufacturing record evaluation was performed for the finished device lot number php808-7665h, and no non-conformances were identified.Additional h3 investigation summary: it was reported that needle pull-off during use.One open box with eleven unopened samples of product code 7665, lot php808 was received for analysis.The complaint sample was not returned for evaluation.During the visual inspection of eleven samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects or damaged or needle were observed.A functional test was performed and the pull force was above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Note: events reported via mw 2210968-2020-07394 (prolene 8833h), 2210968-2020-07395 (monocryl mcp4940h), 2210968-2020-07396 (ethilon 7665h) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|