• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 28sep2020.
 
Event Description
It was reported to philips that the device shutdown during use.The device was reported as being in use at the time of the event on a patient.The initial reporter confirmed that there was no adverse patient impact.
 
Manufacturer Narrative
G4:07dec2020, b4:08dec2020.The customer had a standby ventilator, which was immediately connected to the patient, and there was no delay in therapy.The customer evaluated the device with assistance from a philips remote service engineer (rse).It was determined that the power supply needed to be replaced to return the device to a working condition.The fse replaced the power supply to resolve the reported issue.The device passed performance verification testing.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10593857
MDR Text Key209830126
Report Number2031642-2020-03463
Device Sequence Number1
Product Code MNT
UDI-Device Identifier884838020054
UDI-Public(01)884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-