G4:07dec2020, b4:08dec2020.The customer had a standby ventilator, which was immediately connected to the patient, and there was no delay in therapy.The customer evaluated the device with assistance from a philips remote service engineer (rse).It was determined that the power supply needed to be replaced to return the device to a working condition.The fse replaced the power supply to resolve the reported issue.The device passed performance verification testing.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|