| Model Number |
7213 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/22/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first day of the month of (b)(6) since the only given date of event was (b)(6) 2020.
|
| |
|
Event or Problem Description
|
|
It was reported via user/facility medwatch# (b)(4) that device entrapment and shaft break occurred.The target lesion was located in the left anterior descending artery.A 3.75mm x 15mm nc emerge balloon catheter was advanced for dilatation.However, the balloon was stuck on the wire and while attempting to pull it out of the body, the balloon was noted to have broken off the delivery system.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported via user/facility medwatch# (b)(4) that device entrapment and shaft break occurred.The target lesion was located in the left anterior descending (lad) artery.A 3.75mm x 15mm nc emerge balloon catheter was advanced for dilatation.However, the balloon was stuck on the wire and while attempting to pull it out of the body, the balloon was noted to have broken off the delivery system.No patient complications were reported.It was further reported that the mid lad contained 95% stenosis and 75% in-stent restenosis in distal lad.During removal, femoral access was gained and a 6fr jr4 caheter and a gooseneck catheter were used to successfully remove the balloon tip.
|
| |
|