| Model Number |
70805-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned and an extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle lite meter were reviewed and the dhrs showed the freestyle lite meter passed all tests prior to release.Dhrs (device history review) for the freestyle test strips were reviewed and the dhrs showed the freestyle test strips passed all tests prior to release.Retain testing was performed and all units performed within specification.If the product is returned, the case will be re-opened and a physical investigation will be performed.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 42 mg/dl, 118 mg/dl and 265 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported meter (b)(6) has been returned and investigated with retained test strips.Visual inspection has been performed on the returned meter and no issues were observed.Control solution testing has been performed; all results were within range specification.No malfunction or product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 42 mg/dl, 118 mg/dl and 265 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|