DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter phone number: (b)(6).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Udi: (b)(4).
 
Event Description
It was the reported by the customer as following: the gasket has come off the device after an unknown procedure.No harm to the patient reported.No delay reported.No further information available.Loaner requested.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received at the service center and evaluated.It was reported that the gasket has come off the device after an unknown procedure.Per service manual operational and diagnostic, the reported failure was confirmed.During evaluation, it was identified that the motor was corroded; therefore, it was replaced.The keyboard and cable were in good conditions.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The possible root cause corrosion condition could be related to internal deterioration possible produced by fluid ingress into the motor, as well as lack of maintenance.However, this cannot be conclusively affirmed.A manufacturing record evaluation was performed for the finished device [1843m3950r] number, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: a manufacturing record evaluation was performed for the finished device [1843m3950r] number, and no non-conformances were identified.Device history batch null.Device history review null.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10596762
MDR Text Key208976043
Report Number1221934-2020-02785
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 09/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2020
Initial Date Manufacturer Received 09/14/2020
Initial Date FDA Received09/28/2020
Supplement Dates Manufacturer Received12/11/2020
Supplement Dates FDA Received12/12/2020
Removal/Correction NumberN/A
Patient Sequence Number1
-
-