| Catalog Number |
3910-500-312 |
| Medical Device Problem Code |
Expiration Date Error (2528)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that an expired product was used.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: expired lot used.Probable root cause: design: validated shelf life too short.Application: distribution inefficient, sat too long in inventory before being shipped to customer.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that an expired product was used.
|
| |
|