• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STIMWAVE TECHNOLOGIES INC. FREEDOM NEUROSTIMULATOR; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STIMWAVE TECHNOLOGIES INC. FREEDOM NEUROSTIMULATOR; SPINAL CORD STIMULATOR Back to Search Results
Model Number FR8A-SPR-B0
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The clinical representative confirmed that the implant procedure was performed in a sterile environment, sterile field handling protocols were used, the procedure was completed in accordance with the product instructions for use, and the sterile barriers of all product used were intact prior to implant.On (b)(6) 2020, cr reported covering case and she stated implanting clinician was implanting sand shark over the lead when he severed lead with bevel of sand shark.He removed the severed lead with no issues and re implanted new lead new sand shark.On (b)(6) 2020, cr contacted complaint specialist on the phone to report that the patient had not been in contact.On (b)(6) 2020, complaint specialist, reviewed sterilization and packaging records for the respective product lots, stimwave has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging for lots swo191113.The stimulator was reported to meet product specifications.Manufacturer cannot currently receive the stimulator for analysis.The device was used for treatment of pain.The device cannot be traced as the source of the issue.
 
Event or Problem Description
Stimwave quality has investigated the details regarding a complaint of fractured stimulator reported to stimwave's complaint system on (b)(6) 2020, by clinical representative (cr), in the united states.Stimwave became aware of the issue on (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM NEUROSTIMULATOR
Common Device Name
SPINAL CORD STIMULATOR
Manufacturer (Section D)
STIMWAVE TECHNOLOGIES INC.
1310 park central boulevard s.
pompano beach 33064
MDR Report Key10597408
Report Number3010676138-2020-00134
Device Sequence Number1192007
Product Code GZB
UDI-Device Identifier00818225020648
UDI-Public010081822502064817211101
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170141
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 07/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/01/2021
Device Model NumberFR8A-SPR-B0
Device Lot NumberSWO191113
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2020
Initial Report FDA Received Date09/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-