|
It was reported that revision surgery was performed due to an infection.The patient was opened up, washed out and re-closed again.No parts were explanted.All of the implanted devices were used in treatment.As of today, additional information has been requested for this complaint but have not become available.A review of the complaint history for the bhr head and bhr cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.It was also confirmed that all devices were sterilised.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.Smith and nephew has not received adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.Without further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|