WRIGHTS LANE SYNTHES USA PRODUCTS LLC 14.0MM CANNULATED AWL Back to Search Results
Model Number 03.010.041
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Codes Injury (2348); Foreign Body In Patient (2687)
Date of Event 09/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent intramedullary nail, during the procedure, a piece of the cannulated awl broke off in the greater trochanter.Fragments were generated.The fragment was not able to be removed and remained in the patient.There was no surgical delay.The procedure was successfully completed.Patient status is unknown.This report involves one (1) 14.0mm cannulated awl.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual inspection: the 14.0mm cannulated awl (part #: 03.010.041 and lot #: 1493547) was returned and received at us cq.Upon visual inspection, the tip of the cannulated awl was broken.The broken fragment was returned.A slight discoloration was observed on the device but has no impact on the device functionality.The embedded piece cannot be confirmed as there was no evidence (photos or x-rays) provided showing that the fragments were embedded in the patient.No other issues were identified with the returned device.Device failure/defect identified? yes.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage.Furthermore, the complaint relevant dimensions cannot be checked for dimensional accuracy, because of the design of the device.Document/specification review based on the date of manufacture the following drawings, reflecting the current and manufactured revision were reviewed 14 mm cannulated awl awl tip.Complaint confirmed? yes, the device received was broken.Hence confirming the allegation.Investigation conclusion the complaint condition is confirmed for the 14.0mm cannulated awl (part #: 03.010.041 and lot #: 1493547).However, the embedded device allegation was not confirmed as there were no photos or x-rays provided.There is no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.The potential cause could be due to unintended forces applied to the device.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history part: 03.010.041, lot: 1493547, manufacturing site: hägendorf, release to warehouse date: 10.May.2006.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
14.0MM CANNULATED AWL
Common Device Name
AWL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10599447
Report Number2939274-2020-04360
Device Sequence Number3054118
Product Code HWJ
UDI-Device Identifier10886982067142
UDI-Public(01)10886982067142
Combination Product (Y/N)N
Initial Reporter StateCT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.010.041
Device Catalogue Number03.010.041
Device Lot Number1493547
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Initial Date Received by Manufacturer 09/04/2020
Supplement Date Received by Manufacturer10/16/2020
Initial Report FDA Received Date09/29/2020
Supplement Report FDA Received Date10/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 YR
-
-