| Model Number |
MI1200 SYNCHRONY |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Tissue Breakdown (2681)
|
| Date of Event |
09/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The recipient is planned to be explanted due to electrode extrusion through the tympanic membrane and non-use.The recipient never had benefit from the device due to cystic cochlea deformity and cochlear nerve deficiency.No product deficiency is alleged.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received from the field the device was explanted due to an extrusion of the electrode lead into the external ear canal.Further it is reported that the recipient was a non-user due to no benefit.Reportedly the lack of benefit was caused by device unrelated medical reasons namely cystic cochlea deformity and cochlear nerve deficiency.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
|
| |
|
Event or Problem Description
|
|
The electrode lead extruded through the tympanic membrane.The device was explanted on (b)(6) 2020.
|
| |
|