Model Number DTPB2QQ
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Patient Problems Anxiety (2328); Shock from Patient Lead(s) (3162)
Event Date 09/15/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient received inappropriate high voltage therapy from the cardiac resynchronization therapy defibrillator (crt-d) for atrial fibrillation (af) with rapid ventricular response which was detected as ventricular fibrillation (vf).Additional information reported the right atrial (ra) lead exhibited intermittent undersensing on the treated episodes.The crt-d and ra lead remain in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the patient was given medication.It was also noted that the patient experienced psychological distress and prolonged hospitalization due to the inappropriate shocks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT HF QUAD CRT-D MRI SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key10600580
MDR Text Key208982450
Report Number9614453-2020-03051
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00763000178178
UDI-Public00763000178178
Combination Product (y/n)N
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/14/2021
Device Model NumberDTPB2QQ
Device Catalogue NumberDTPB2QQ
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/15/2020
Initial Date FDA Received09/29/2020
Supplement Dates Manufacturer Received10/05/2020
Supplement Dates FDA Received10/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
5076-45 LEAD, 429888 LEAD, 6935M55 LEAD; 5076-45 LEAD, 429888 LEAD, 6935M55 LEAD
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age80 YR
-
-