• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB TRAY SZ4 LMRL; EARLY INTERVENTION : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB TRAY SZ4 LMRL; EARLY INTERVENTION : KNEE TIBIAL TRAY Back to Search Results
Model Number 1024-51-400
Medical Device Problem Codes Loss of or Failure to Bond (1068); Malposition of Device (2616)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 09/16/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient underwent primary lm unicompartmental knee replacement on (b)(6) 2019 where a sigma partial was utilised ((b)(6)).Patient has reported that the knee has never felt totally good.Patient has reported ongoing pain.X-rays indicate a possible loose tibial component with possible sub-optimal positioning at time of index surgery.A decision was made to revise the sigma partial to a total knee on (b)(6) 2020 ((b)(6) hospital).On opening the knee, the tibial tray was not well fixed.All components were removed and a stemmed tibial component (sleeve and stem) implanted on the tibial side and a primary attune ps femur on the femoral side.Patient status/ outcome / consequences: unknown, patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)? pain and possible tibial loosening, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown, is the patient part of a clinical study: no, (b)(4).Device property of: none, device in possession of: none, (b)(4).Device property of: none, device in possession of: none, (b)(4).Device property of: none, device in possession of: none.By checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : no deviations or non-conformance's were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI TIB TRAY SZ4 LMRL
Common Device Name
EARLY INTERVENTION : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10601071
Report Number1818910-2020-21122
Device Sequence Number3321703
Product Code HRY
UDI-Device Identifier10603295002345
UDI-Public10603295002345
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K070267
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1024-51-400
Device Catalogue Number102451400
Device Lot NumberHC1716
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/16/2020
Supplement Date Received by Manufacturer12/09/2020
Initial Report FDA Received Date09/29/2020
Supplement Report FDA Received Date12/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMPETITOR - PALACOS.; SIGMA HP UNI FEMORAL SZ4 LM/RL.; SIGMA HP UNI INS SZ4 8MM LM/RL.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-