AESCULAP AG AS VEGA PS TIBIAL PLATEAU CEMENTED T0; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NX049Z
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Nx049z: device name as vega ps tibial plateau cemented t0, serial number n/a, batch number unknown, udi device identifier (b)(4), udi production identifier unknown, basic udi-di n/a, unit of use udi-di (b)(4), manufacturing date unknown.Ref.Code device name batch: nx009z as vega ps femoral comp.Cemented f4n l 52137600.Nx100 vega ps gliding surface t0/0+ 10mm 51947691.Nn260p plug f/tibial plateau 52119154.Nx042 patella 3-pegs p2 52120719.Investigation: no product at hand.Therefore an investigation at the product is not possible.Pictorial documentation: there are no pictures available.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.There are 3 similar complaints against the same lot number 52137600.(2 as involved components).Explanation and rationale: in the light of the little information received and due the circumstance that we did not received the complained devices it is not possible to determine a root cause for the mentioned failure.Corrective action: for this topic (vega loosening) a product safety case (psc) was initiated.
 
Event or Problem Description
It was reported that there was an issue with as vega knee.As a result of having the product implanted the patient has experienced knee pain, swelling, difficulty walking, and loosening and instability of the implant.The primary surgery occurred on (b)(6) 2015 and the revision surgery occurred on (b)(6) 2016.All available information has been provided at this time.If additional information becomes available the complaint will be updated accordingly.The adverse event / malfunction is filed under reference (b)(4).Associated medwatches: 2916714-2020-00411, 2916714-2020-00412, 2916714-2020-00414, 2916714-2020-00415.Involved components: nx009z (ps femur cemented f4n lt).Nx100 (ps pe insert t0/t0+, 10mm).Nx049z (ps tibia cemented t0).Nn260p (peek plug f/ tibia).Nx042 (universal patella p2).The cement used is unidentified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AS VEGA PS TIBIAL PLATEAU CEMENTED T0
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10601799
Report Number2916714-2020-00413
Device Sequence Number3273149
Product Code HRY
UDI-Device Identifier04046963810337
UDI-Public4046963810337
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNX049Z
Device Catalogue NumberNX049Z
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/15/2018
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date09/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-