• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE ADJ TIB STYLUS; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE ADJ TIB STYLUS; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES Back to Search Results
Model Number 2544-00-007
Device Problems Material Twisted/Bent (2981); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/16/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
During training with surgeons earlier this week, wednesday 16 september at 12.00 it was noted that attune tibia stylus art nr 254400007 lot nw207641 was not functioning ¿ looks bent and is not able to put in to the cuttingblocks without getting stuck.This is a new instrument, never been used before, has been cleaned and sterilized by the hospital (b)(6).Delivered to the hospital 14 september.The instrument has been replaced with a new one.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device did not confirm the reported event.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE ADJ TIB STYLUS
Type of Device
ATTUNE INSTRUMENTS : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10602680
MDR Text Key209391829
Report Number1818910-2020-21172
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295129790
UDI-Public10603295129790
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2544-00-007
Device Catalogue Number254400007
Device Lot NumberNW207641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2020
Date Manufacturer Received10/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-