• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 5-6 9MM LT; PROSTH,KNEE,PATFEMTIB,SEMCONSTR,CEMNT,POL/METL/POL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 5-6 9MM LT; PROSTH,KNEE,PATFEMTIB,SEMCONSTR,CEMNT,POL/METL/POL Back to Search Results
Model Number 74035261
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during tka surgery, the journey ii bcs art insert trial size 5-6 9mm left broke outside the patient.The procedure was finished using a smith and nephew back up device, with no surgical delay and no injury to the patient.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).Results of investigation: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured, rendering the device inoperable.The device was manufactured in 2016 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS ART ISRT TRL SZ 5-6 9MM LT
Common Device Name
PROSTH,KNEE,PATFEMTIB,SEMCONSTR,CEMNT,POL/METL/POL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10602834
Report Number1020279-2020-04933
Device Sequence Number3054115
Product Code JWH
UDI-Device Identifier00885556157039
UDI-Public00885556157039
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74035261
Device Catalogue Number74035261
Device Lot Number16KAP3496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Initial Date Received by Manufacturer 09/01/2020
Supplement Date Received by Manufacturer11/19/2020
Initial Report FDA Received Date09/29/2020
Supplement Report FDA Received Date11/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-