Patient Problem
No Consequences Or Impact To Patient (2199)
Event Type
Malfunction
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.
The reported problem was confirmed.
The unit was tested and the lock latch mechanism found to cause a distorted image when manipulating the test scope.
In addition, the eject switch on the pcb was getting stuck intermittently.
Event Description
A user facility reported to olympus that the image produced was distorted.
The problem, as reported to olympus, occurred during a procedure.
There was no patient injury or harm, associated with the problem, reported to olympus.