Model Number CI-1601-05 |
Device Problem
Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Therapeutic Response, Decreased (2271)
|
Event Date 08/26/2020 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced decreased performance.The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the top and bottom cover.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical test performed.Advanced electrical testing performed on the device revealed lower than typical range on some of the electrodes.It is the opinion of advanced bionics explant review board that some of these electrode test results should be considered as failures.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|