|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.Broken devices can be a result of, but not limited to, manufacturing, mishandling, or external force applied to the device (tapping indeflator to clear air bubbles, strong twisting during inflation/deflation).It is possible that inadvertent mishandling and/or an excessive amount of force was applied to the device while attempting to deflate resulting in the reported break; however this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|