RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.(b)(6)2020.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
The customer reported that the battery is not charging.The customer contacted product support and was advised to try swapping another power management board from another unit.The biomed states that the battery is the issue.The biomed states that battery manufacture date is 2017.The customer reported that the unit was not in use on a patient.The issue was discovered when they turned on the unit for set up.No delay was reported.The biomedical engineer replaced the battery to address the reported issue.
|
|