Model Number D132704
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Death (1802); Vascular Dissection (3160); No Code Available (3191)
Date of Event 08/31/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
(b)(4).The bwi product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(4).
 
Event or Problem Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure on (b)(6) 2020 with a thermocool® smart touch¿ bi-directional navigation catheter and vessel dissection (requiring surgical intervention) and death.Initially it was reported that during the procedure, the catheter was bent and was not deflecting properly to the f curve.The catheter was replaced, and the issue resolved.No patient consequence was reported.The reported deflection issue was assessed as not mdr reportable.Insufficient deflection or not relaxing completely (without stuck in the deflected position) would cause intraprocedural delay.The potential risk that it could cause or contribute to a death or serious injury was remote.Additional information was received on (b)(6) 2020.Post-procedure while the patient was in recovery, the patient had developed some internal bleeding.It was believed the patient had an aortic dissection; however, it is unknown how the dissection was discovered or confirmed.It had been noted that the patient had been stable throughout the procedure.Additional surgeries were required.It was reported that the patient died on (b)(6) 2020.No further information is available.Physician causality opinion was not provided.Any product malfunction or serious injury that results in patient¿s death is to be considered serious and mdr-reportable.The awareness date for the report table adverse event is (b)(6) 2020.
 
Additional Manufacturer Narrative
On 10/1/2020, the product investigation was completed.It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure on (b)(6) 2020 with a thermocool® smart touch¿ bi-directional navigation catheter and vessel dissection (requiring surgical intervention) and death.Initially it was reported that during the procedure, the catheter was bent and was not deflecting properly to the f curve.The catheter was replaced, and the issue resolved.No patient consequence was reported.Device evaluation details: the device was visually inspected, and it was found in good conditions.The magnetic, temperature, and force features were tested, and no issues were observed.In addition, the catheter was deflecting and irrigating correctly.No malfunctions were observed during the product analysis.A manufacturing record evaluation was performed, and no internal actions related to the reported complaint were identified.The catheter passed all specifications.The root cause of the adverse event remains unknown.The instructions for use (ifu) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation or tamponade.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL SMART TOUCH BI-DIRECTIONAL NAVIGATION CATHETER
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
33 technology drive
irvine CA 92618
MDR Report Key10603943
Report Number2029046-2020-01360
Device Sequence Number2803369
Product Code LPB
UDI-Device Identifier10846835009194
UDI-Public10846835009194
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P030031/S053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/22/2021
Device Model NumberD132704
Device Catalogue NumberD132704
Device Lot Number30373117M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Initial Date Received by Manufacturer 09/02/2020
Supplement Date Received by Manufacturer10/01/2020
Initial Report FDA Received Date09/29/2020
Supplement Report FDA Received Date10/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Required Intervention;
-
-