| Model Number |
37302 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported the adhesive of the male external catheter changed and was now sticking too well.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed since product meets specifications.Visual evaluation noted 42 unopened spirit mecs were received.No obvious defects were noted.Further testing required.Adhesive peel test was performed.Samples were tested.Samples were cut to the required width (0.70" +/- 0.05").Adhesive peel strength was found to be within specification (0.60-2.10lbf).The device history record review was not required as the reported event was unconfirmed.The instructions for use were found adequate and state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." corrections: d4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported the adhesive of the male external catheter changed and was now sticking too well.Per additional information received on 30sep2020, it stated that the male external catheter's adhesive has changed to be stronger and also stated it was impossible to get off.
|
| |
|
Search Alerts/Recalls
|