• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 3-4 LT 9MM; PROSTH,KNEE,PATFEMTIB,SEMCONSTR,CEMNT,POL/METL/POL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 3-4 LT 9MM; PROSTH,KNEE,PATFEMTIB,SEMCONSTR,CEMNT,POL/METL/POL Back to Search Results
Model Number 74027241
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Date of Event 09/01/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery of journey ii bcs xlpe articular insert sz 3-4 lt 9mm was performed due to patient experienced pain and infection.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms this complaint from japan reports that a revision was performed secondary to pain and infection.The doctor says the devices were not defective.The date of the primary surgery is not known.Smith and nephew has not received the device/adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Infection, a potential complication associated with any surgery, can occur and possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS XLPE ART ISRT SZ 3-4 LT 9MM
Common Device Name
PROSTH,KNEE,PATFEMTIB,SEMCONSTR,CEMNT,POL/METL/POL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10604290
Report Number1020279-2020-04945
Device Sequence Number2941355
Product Code JWH
UDI-Device Identifier00885556177617
UDI-Public00885556177617
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74027241
Device Catalogue Number74027241
Device Lot Number20BM07970
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/01/2020
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date06/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-