Model Number 174213
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, pre-operatively of a rectal procedure, when the device was opened, the flexible part at the tip was already broken.Another device was used to complete the case.There was no patient involvement.
 
Additional Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the plastic cable sheath had shearing damage.Functional evaluation found that the jaw rotation worked without difficulty; re lease lever, ratchet button and jaw toggle functioned properly; and, jaw articulated, opened and closed without difficulty.It was reported that the sheath was broken.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.Consult the medical literature relative to complications, hazards and techniques, prior to performing endoscopic procedures.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO DISSECT
Common Device Name
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key10606363
Report Number2647580-2020-03188
Device Sequence Number16746613
Product Code GET
UDI-Device Identifier10884523000665
UDI-Public10884523000665
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K914753
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number174213
Device Catalogue Number174213
Device Lot NumberP8K1474X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Initial Date Received by Manufacturer 09/14/2020
Supplement Date Received by Manufacturer10/28/2020
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date11/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-