Model Number 10619
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/21/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: synergy ous mr 2.50 x 20 mm stent delivery system was returned for analysis.A visual examination of the stent found signs of stent damage.Stent struts from the distal stent region were noted to be lifted form their crimped position and pulled in a proximal direction.The undamaged stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube shaft found multiple kinks along the length of the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on 08 sep 2020.It was reported that crossing difficulties were encountered.The target lesion was located in the severely tortuous and severely calcified left anterior descending artery.A 2.50 x 20mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with a different device.There were no patient complications reported and the patient was in good condition.However, returned device analysis revealed stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10606974
MDR Text Key209186219
Report Number2134265-2020-13541
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2020
Device Model Number10619
Device Catalogue Number10619
Device Lot Number0024014695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/12/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/08/2020
Initial Date FDA Received09/30/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-