MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 32IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR +4 10D 32IDX52OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1219-32-152
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Health Effect - Clinical Codes Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 09/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
A patient was treated with a hip tep implantation in 2017 and now complained about sudden restriction of range of movement and a squeaking sound in the left hip joint.Intra op x-ray shows an eccentric placement of the hip head and a subluxated inlay with expected pre-damage.X-ray post op inconspicuous.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed a disassociation event.Error in materials or manufacture was not identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: dhr reviewed and the product had one hold to correct paperwork only, no anomalies or deviations for the product was found.Corrected: h3 and h6 by removing pec (implant noise).
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Aspiration of left hip occurred on (b)(6) 2020: negative for infection.Revision operative notes (b)(6) 2020 indicate the patient received a left total hip revision due to inability to ambulate, inlay damage and subluxation from the cup (disassociation).Upon entering the joint, it is indicated that there were 2 chips that came off of the inlay, these were removed from the joint.The head and liner were exchanged.The surgery was completed.These revision notes go on to indicate that after a review of the prosthetic stickers when preparing the surgery report on the following morning revealed that the wrong inlay (duraloc) had been inserted into the cup (pinnacle).The patient was informed and a revision was planned.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Patient code: removed code for absence of treatment and replaced with no code available (3191) to capture device revision or replacement.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 32IDX52OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10606992
Report Number1818910-2020-21231
Device Sequence Number2793802
Product Code LPH
UDI-Device Identifier10603295014058
UDI-Public10603295014058
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number1219-32-152
Device Catalogue Number121932152
Device Lot NumberH74569
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2020
Initial Date Received by Manufacturer 09/17/2020
Supplement Date Received by Manufacturer10/28/2020
11/08/2020
11/17/2020
12/08/2020
Initial Report FDA Received Date09/30/2020
Supplement Report FDA Received Date11/13/2020
11/17/2020
11/30/2020
12/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORAIL2 NON COL HO SIZE 10; HIP BALL BIOLOX 32 +9 12/14; PINNACLE SECTOR II CUP 52MM; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight72
-
-