| Model Number |
24974 |
| Medical Device Problem Codes |
Break (1069); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/08/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device almost broke.A mach 1 catheter guide was selected for use.The intended lesion was located in the right carotid artery.During the procedure, the physician had difficulty moving the device inside the arteries.When moving the device, it almost broke inside the patient due to the weakness in the middle of the catheter.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.The catheter shaft was inspected for damage.The catheter shaft showed 1 kink located 48cm from the tip.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that the device almost broke.A mach 1 catheter guide was selected for use.The intended lesion was located in the right carotid artery.During the procedure, the physician had difficulty moving the device inside the arteries.When moving the device, it almost broke inside the patient due to the weakness in the middle of the catheter.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient was stable post procedure.
|
| |
|