Model Number 72200561S
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2020
Event Type  malfunction  
Event Description
It was reported that during a hip arthroscopy the camera head had a flicker image.The procedure was successfully completed using a back up device.A delay of 5 minutes was reported.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of flickering could not be reproduced.Product passed functional testing and 2 hour burn-in.Flickering did not occur during functional testing and burn-in.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC CAMERA HEAD NON-AC HD560H
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10607298
MDR Text Key209233515
Report Number1643264-2020-01348
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier03596010606808
UDI-Public03596010606808
Combination Product (y/n)N
PMA/PMN Number
K070266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200561S
Device Catalogue Number72200561S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Date Manufacturer Received09/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-